CRP Rapid Test - C-Reactive Protein Cartridges
CRP Rapid Test - C-Reactive Protein Cartridges
CRP - C-Reactive Protein test, test kit for whole blood, fingertip blood, serum and plasma. For use with the POCT SmartTester.
Features
Easy to Use CRP rapid test
CRP test provides rapid results to POCT with a 10 microlitre blood sample, with a reading time of only 1.5 minutes.
The CRP Rapid lateral flow test is an immunoassay designed for the quantitative determination of C- reactive protein (CRP) concentration in whole blood, fingertip blood, serum and plasma. For professional in vitro diagnostic use only.
CRP is virtually absent from the serum of a healthy person, and is the most sensitive acute phase protein produced in the liver as a response to microbial infection or tissue injury.
The measurement of CRP in the blood is valuable in detecting inflammation , infection or in monitoring disease in acute conditions.
The rapid diagnostic CRP test uses an anti-human CRP monoclonal antibody conjugated with colloidal gold and another anti-human CRP monoclonal antibody coated on the test line.
Sample: |
Serum, plasma, whole blood, fingertip |
|
Anticoagulant: |
Heparin, sodium citrate, EDTA. |
|
Sample Volume: |
10ul |
|
Reaction Time: |
1.5 minutes |
|
Storage: |
4 - 30°C |
|
Shelf life: |
Up to 24 months |
Kit Contents
Contents
- 25 CRP test cartridge
- 25 Dilution buffer
- 1 Lot code cartridge
- 25 Disposable capillary pipet
- 1 IFU
Test Research
Method was compared to the Beckman Coulter AU 5800 analyser according to CLSI-EP09-A3 guideline.
- Range 0.5 – 200mg/L
- Limit of Detection 0.5mg/L
- Linearity R = 0.9887 [3]
- Method Comparison to Beckman AU5800
[3] Clinical and Laboratory Standards Institute, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach, Approved Guideline. NCCLS Document EP6-A,” NCCLS Doc. EP6-A, vol. 23, no. 16, pp. 1–50, 2003.
A total of 150 samples were used for this evaluation with CRP levels throughout the measuring range.
The results from the linear regression analysis of the comparison data are shown below.
Variants and Related Products
Variants and Related Products
Order Code | Contents | Size |
STR-9911 | SmartTester Reader | 1 device |
RTC-9901-1 | CRP Rapid Test Cartridges | 25 tests |
RTC-9913-1 | hCG Rapid Test Cartridges | 25 tests |
RTC-9902-1 | D-Dimer Rapid Test Cartridges | 25 tests |
RTC-9911-1 | Microalbumin Cartridges | 25 tests |
RTC-9912-1 | TSH Cartridges | 25 tests |
RTC-9903-1 | Cardiac Troponin l (cTnI) Rapid Test cartridges | 25 tests |